Once the electrical connectors and sample tubes have been attached, check the
unit for leaks. Remember not to run the motor for more then a few seconds
without liquid in the tank.
Cleaning the cell windows
Cleaning the system by rinsing through with fresh dispersant a couple of times is
usually sufficient to prepare for a new measurement. Over time you will notice
you cannot achieve such a good background measurement. That is the time to
clean the cell windows.
The cell windows are held in place by a metal retaining ring that can be
unscrewed by using the window tool. The cell windows are sealed by “O” rings.
$
To remove a cell window from the Mastersizer S or X:
.
Make sure that the cell has been drained of dispersant and remove the cell
assembly from the Mastersizer bulkhead by releasing the locking lever.
.
Disconnect the two sample tubes from the cell.
.
Locate the two pegs on the window tool
À
into the two holes on the retain-
ing ring
Á
.
.
Rotate the window tool anti-clockwise and remove the ring
Á
.
.
Tip the window
Ã
out of the cell onto a clean paper towel.
.
If the “O” ring
Â
did not come out with the window then carefully remove
it with a pair of tweezers.
The window assembly is shown in the diagram below.
5
5
1
2
4
3
ILL2983
C H A P T E R 9
Q S
S M A L L
V O L U M E
S A M P L E
D I S P E R S I O N
U N I T
P a g e 9 . 5
Summary of Contents for Mastersizer X
Page 7: ...C H A P T E R 1 Introduction to this manual Introduction to this manual...
Page 8: ......
Page 15: ...C H A P T E R 2 Health and safety Health and safety...
Page 16: ......
Page 23: ...C H A P T E R 3 Installation Installation...
Page 24: ......
Page 35: ...C H A P T E R 4 Getting to know your Dispersion Unit Getting to know your Dispersion Unit...
Page 36: ......
Page 43: ...C H A P T E R 5 How the Accessory works How the Accessory works...
Page 44: ......
Page 50: ......
Page 53: ...C H A P T E R 7 Creating changing SOP s Creating changing SOP s...
Page 54: ......
Page 61: ...C H A P T E R 8 Sample preparation Sample preparation...
Page 62: ......
Page 65: ...C H A P T E R 9 Maintenance Maintenance...
Page 66: ......
Page 75: ...A P P E N D I X A Specification Specification...
Page 76: ......
Page 79: ...A P P E N D I X B Chemical compatibility Chemical compatibility...
Page 80: ......
Page 83: ...A P P E N D I X C Simple fault finding Simple fault finding...
Page 84: ......
Page 87: ...A P P E N D I X D Installing the Qspec MS1 macro Installing the Qspec MS1 macro...
Page 88: ......
Page 93: ...A P P E N D I X E EMC performance EMC performance...
Page 94: ......
Page 97: ...Index Index...
Page 98: ......