
51
is 37
The device will start heating from -5 to 100
when
temperature arrival 100
and then to start countdown from 30 min to 0 min.
The device will alarm “beep” while T
3
100 and 0 min. T
3
value is flashing,
and then use
key and
key to enter the real temperature as
thermometer. Press
key to store the updated value.
(Set a calibration target temperature) (Start calibration)
(Key in the same value as thermometer) (Arrival the target temperature)
6. Please wait for few more minutes that microprocessor will auto adjust
temperature until display value is the same as thermometer.
DOMINIQUE DUTSCHER SAS
Summary of Contents for MC-0203
Page 19: ...18 D O M I N I Q U E D U T S C H E R S A S...
Page 35: ...34 Step6 Complete the installation program D O M I N I Q U E D U T S C H E R S A S...
Page 55: ...54 MEMO D O M I N I Q U E D U T S C H E R S A S...
Page 56: ...55 MEMO D O M I N I Q U E D U T S C H E R S A S...
Page 57: ...56 MEMO D O M I N I Q U E D U T S C H E R S A S...
Page 58: ...57 MEMO D O M I N I Q U E D U T S C H E R S A S...