29
Note (1):
When running Annealing Program Mode, the current temperature of block and
the step settings will show on the screen
There are two situations for running time:
Case 1: set time is zero. Once the current block temperature reaches the set
temperature, it will go to the next step immediately.
Case 2: set time is great then zero. Once the current block temperature reach
the set temperature, the running time starts counting down to zero and stop
temperature control.
Press
key to display the set temperature and time. It will be kept on
screen for 5 seconds or just press
key again to go back monitor
screen immediately.
Note (2):
The final step in Annealing Mode step parameters will run 999
minutes unless stop by user.