25
Note (2):
When running Programmable Mode, the current temperature of block and the
step settings will show on the screen.
You can set up all 4 steps, or, for example, if you only need 3 steps (Step0 –
Step2) in the program, set the running time in Step3 at 0 (zero). Then when the
program finishes Step2, it will return to the first step or stop the temperature
control (i.e., gradually return to the ambient temperature), depending on the
numbers of cycles you set. The following charts are examples:
ambient
temp.
Programmable Mode
(Cycle:1)
The temperature in Step3 can be set at any value. When the running time in one
of steps is set at 0, the program will only run the previous steps.
And in this example, the running time in Step3 is 0, as a result, the system will
recognize only 3 steps in this program.