Operating Instructions MI 34 / MI 34 H
GEBAmi34e.11a.docx
47
850 232/14 08/11
13 Care and maintenance of the instrument
Disconnect the power plug before cleaning!
To clean the instrument, probe, contralateral receiver and other accessories
use a soft cloth dampened with a little warm soapy water or washing-up
liquid; no alcohol or spirits should be used. During cleaning, please ensure
that no liquid runs into the switches, level control or probe openings.
Please use a new eartip for each patient. Always use eartips from MAICO or
Sanibel. Each eartip should only be used one time.
13.1 Cleaning of Probe Tip
In order to secure correct impedance measurements it is important to make
sure that the probe system is kept clean at all times. Therefore please follow
the below illustrated instruction on how to remove e.g. cerumen from the
small acoustic and air pressure channels of the probe tip. For the MI 34 two
different probe systems exist; the Screening Probe System and the
Diagnostic Probe System. The two different probe systems can be seen in
the below picture:
Figure 38
: