Operating Instructions MI 34 / MI 34 H
GEBAmi34e.11a.docx
14
850 232/14 08/11
Figure 8 MI 34 probe with additional screening
probe tip
Â
E
M
E
N
E
O
E
P
E
Q
E
R E
S
E
X
3.8 Getting familiar with the probe
The probe of the MI 34 is shown
in figure 8. The probe head is
adjustable in three steps (0°, 60°
und 80°). It is adjusted by
releasing the fixation screw
(Figure 9
¾
) at the bottom of
the probe a few turns, using a
coin or a screw driver. Adjust the
probe head
º
by pulling it into
the required position until it
rests. To do this hold the probe
handle
¸
with the other hand. After it is set to the required position fasten
the fixation screw
¾
again.
Notice! To avoid damages on the sensitive measuring
equipment, only bent the probe toward fixation screw!
Transforming the Probe from Handheld to Clinical and Reverse:
To exchange the probe insert press the release button
¹
of the probe with
a tool or a pen. Remove the screening probe insert.
Insert the diagnostic probe insert
¿
into the probe head
º
. Please
recognize the right position of the connector of the diagnostic probe insert
¿
. Press the diagnostic probe insert
¿
into the probe head
º
until it
fastens.
Note! The squared hole in the probe connection should face in the same
direction as the push-button.
Control lights and Display
The probe button
·
can be used to select the required test ear. The colour
of the control light
¹
changes accordingly to red (right ear) or blue (left
ear). If selected in the setup menu pressing the probe button
·
during
¾
Figure 9