Operating Instructions MI 34 / MI 34 H
GEBAmi34e.11a.docx
21
850 232/14 08/11
0.94 ml
0.81 ml
- 37 daPa
32 daPa
ml
3
2
1
0
-600
-300
300 daPa
Right
Ear Volume
Compliance
Pressure
Gradient
0.94 ml
0.81 ml
-37 daPa
31 daPa
Tympanogram
Remarks:
Examiner:
Name:
Id No.:
Date:
MAICO MI 34
Figure 16 – Printout of Tympanogram
4.6 How to print out the test result
After the end of a test you can print out
the results for your records by pressing the
PRINT button
Ã
. The quiet thermal printer
prints out the example used in the previous
paragraph 4.5 in only 6 seconds.
While the printer is working no key action
is possible and the probe is inactive. Figure
16 shows the printout:
Id No.:
Here you can put down the
patients social Id number.
Date:
Here the actual test date can be
stated.
Name:
Here you can put down the
patients name.
Examiner:
State here the reference for the
test person.
Remarks:
Additional information about
the test or patient can be stated here.
All other values and the Tympanogram
correspond to those you have seen on the
LCD display and which were explained on
the previous page under 4.5.
The “intelligent” printer control helps you to save paper. It will only print
out what has really been measured. So the printout of the reflex
frequencies misses in the example above because only the Tympanogram
was measured.
If you have saved two Tympanograms (for example for both the left and the
right ear) both are printed out side by side.
You can produce as many printouts as you want by pressing several times
the PRINT button
Ã
.
4.7 How to delete the test results
By pressing the
R/L
-key
Ä
longer the measurement memory will be
deleted. On the LCD-display
Ç
the message “Delete all Data?” occurs.
Press the ENTER button
Ï
to delete all patient data. Then the LCD display
shows an empty measurement screen. If you press the MENU button
Ê
you
return to the measurement screen without deleting the measurement data.