Operating Instructions MI 24, MI 24h
46
16
Technical Data and Accessories
Impedance measurement:
Type:
Clas 2 acc. to ANSI S3.39
Tympanometer:
Test frequency:
226 Hz ± 1%
Test level:
85 dB
SPL
in 2 cm
3
Pressure range:
+200 to -400 daPa
Volume range:
0.1 to 6.0 ml
Accuracy:
± 5 % or ± 10 daPa
Test time:
Less then 3 seconds typical
Compliance range:
0.2 to 3.0 ml
Reflex measurement:
Test frequencies:
500 Hz, 1 kHz, 2 kHz, 4 kHz ± 2%
Test method:
ipsilateral, contralateral (optional)
Test level ipsilateral:
70 dB
HL
to 105 dB
HL
(High freq. option)
(for 4 kHz to 105 dB
HL
)
Test level contralat.:
70 dB
HL
to 110 dB
HL
(High freq. option)
(for 4 kHz to 105 dB
HL
)
Attack/release time:
Typical 10 ms
ON/OFF Ratio:
Greater than 70 dB
Pressure at test:
Pressure @ max. compliance
General:
Memory:
Storage of two complete test result sets
LCD-display:
Graphical display of the tympanograms and the reflex
curves
Numeric display of max. compliance, pressure at max.
Compliance, canal volume, gradient and reflex
thresholds
Printer:
Thermal printer, paper roll width 110 mm
Printing time:
Between 4 seconds (one tympanogram) and 12
seconds (tympanogram and reflex for both ears)
Voltage:
96 - 240 V / 50/60 Hz
Power consumption:
approximately 65 VA
Summary of Contents for MI 24
Page 5: ...Operating Instructions MI 24 MI 24h 3 NOTE 1 02 mm H2O 1 0 daPa...
Page 8: ...Operating Instructions MI 24 MI 24h 6...
Page 11: ...Operating Instructions MI 24 MI 24h 9 24...
Page 47: ...Operating Instructions MI 24 MI 24h 45 Tested by Date...
Page 49: ...Operating Instructions MI 24 MI 24h 47...
Page 52: ...Operating Instructions MI 24 MI 24h 50...
Page 53: ...1162 0321 REV E 07 07...