![Maico MI 24 Operating Instructions Manual Download Page 20](http://html1.mh-extra.com/html/maico/mi-24/mi-24_operating-instructions-manual_3704469020.webp)
Operating Instructions MI 24, MI 24h
18
4.5
How to evaluate the tympanogram display
After
having
completed
a
measurement you can see the
results on the LCD.
On the left side of the display
you see a graph of the
tympanogram.
The
area
surrounded by the box is valid
for “normal” tympanograms.
You can change the area or turn
it off.
For
details
see
chapter 10
“Individual Setup of the
MI 24.”
In the top-middle of the LCD,
the words
Right
or
Left
indicate
the ear chosen at the moment.
Tympanometer
(at the top right) - indicates that the reflex measurement has been
switched off.
In the boxes to the right the following test measurements are displayed:
Ear Volume
- indicates the volume of the section of the auditory canal between the
ear tip and the eardrum in ml (in the example 0.94 ml).
Compliance
- indicates the maximum value of the compliance from the
tympanogram in ml (in the example 0.81 ml).
Pressure
- indicates the pressure with the highest measured compliance (in the
example -37 daPa).
Gradient
- calculations are reported as the tympanogram width at half of peak
compliance expressed in daPa (in the example 32 daPa).
Status
READY
Ear Volume
Compliance
Pressure
Gradient
Impedance
Tympanometer
Right
ml
3
2
1
0
-600
-300
300 daPa
Scale
0.94 ml
0.81 ml
- 37 daPa
32 daPa
Figure 14
Display of a normal tympanogram
Summary of Contents for MI 24
Page 5: ...Operating Instructions MI 24 MI 24h 3 NOTE 1 02 mm H2O 1 0 daPa...
Page 8: ...Operating Instructions MI 24 MI 24h 6...
Page 11: ...Operating Instructions MI 24 MI 24h 9 24...
Page 47: ...Operating Instructions MI 24 MI 24h 45 Tested by Date...
Page 49: ...Operating Instructions MI 24 MI 24h 47...
Page 52: ...Operating Instructions MI 24 MI 24h 50...
Page 53: ...1162 0321 REV E 07 07...