Operating Instructions MI 24, MI 24h
14
If the probe tip is clogged, you can remove it by opening the fixation ring . After
cleaning the probe tip or attaching a new one, the tip must be fixed again by
fastening the fixation ring.
3.7
Choose an appropriate ear tip
Choose an ear tip of the appropriate size from the ear tip set. Put the ear tip tightly
on the probe tip. The probe tip should end near the end of the ear tip. It should not
be inserted more than about 1 mm inside the ear tip or just protrude out of the ear
tip (see Figure 9).
By choosing an appropriate ear tip and placing it correctly on the probe you create
the basic conditions for measurements without problems and mistakes.
Now all preparations are concluded and you can start the impedance and reflex
measurement. Please read the following chapters.
min 0 mm
max 1 mm
Figure 9
Proper placement
of ear tip