Operating Instructions MI 24, MI 24h
24
6.4
Taking the ipsilateral measurement
Carry out the measurement as
described in chapter 4.4 “Recording
the Tympanogram.” The Stapedius
reflex
is
measured
after
the
measurement of the tympanogram.
During the measurement of the
Stapedius reflex the change of the
compliance is represented in real
time on the LCD. When the test is
finished the curves for the changes
of compliance at 500 Hz, 1000 Hz,
2000 Hz and 4000 Hz are shown in
four separate graphs at the bottom
of the measurement screen (see
Figure 20). Below each curve is the
test level where a Stapedius reflex
was registered automatically. This is
indicated by a
Pass
below the
frequency. If no reflex was detected, a
Fail
is reported and the maximum level is
shown.
You can judge this by watching the real time graph if you have a real Stapedius reflex
or only disturbance and artifacts. The lower dotted zero-line of a graph indicates the
measured compliance without a test sound. All the positive or negative changes of
compliance are shown as deviation from the zero-line. If a Stapedius reflex occurs,
the compliance increases and the curve rises. The box which appears during the test
symbolizes the threshold at which the
MI 24
accepts a change of compliance as a
valid Stapedius reflex
.
Status
READY
Ear Volume
Compliance
Pressure
Gradient
Impedance
Ipsi
Right
ml
3
2
1
0
-600
-300
300 daPa
1
100
500 Hz I
4
95
4000 Hz I
3
100
2000 Hz I
2
100
1000 Hz I
AUTO
dB
Scale
0.94 ml
0.81 ml
- 37 daPa
32 daPa
PASS
PASS
PASS
PASS
Figure 20
Example of a normal tympanogram
with ipsilateral reflex results
Summary of Contents for MI 24
Page 5: ...Operating Instructions MI 24 MI 24h 3 NOTE 1 02 mm H2O 1 0 daPa...
Page 8: ...Operating Instructions MI 24 MI 24h 6...
Page 11: ...Operating Instructions MI 24 MI 24h 9 24...
Page 47: ...Operating Instructions MI 24 MI 24h 45 Tested by Date...
Page 49: ...Operating Instructions MI 24 MI 24h 47...
Page 52: ...Operating Instructions MI 24 MI 24h 50...
Page 53: ...1162 0321 REV E 07 07...