![Maico MI 24 Operating Instructions Manual Download Page 29](http://html1.mh-extra.com/html/maico/mi-24/mi-24_operating-instructions-manual_3704469029.webp)
Operating Instructions MI 24, MI 24h
27
7
Interpreting Test Results
7.1
Understanding the printout
The printout contains the following
information:
Ear volume
Compliance
Pressure
Gradient
Reflex Test Results (
Pass
,
Fail
)
and
ipsi
,
contra
or
tympano-
gram
(depending on the test
you have performed).
This information provides the data
you need to interpret the test
results.
A graph of the tympanogram is
provided (Figure 24) to assist you in
a visual interpretation of the test.
This graph is a representation of
the relative mobility of the middle
ear system. The horizontal axis
shows the changes in air pressure
and the resulting mobility of the
system. The compliance is recorded
on the vertical axis. This mobility is
expressed as a change in the
volume of the ear canal in ml.
The reflex is shown in up to four
graphs with time on the horizontal
axis and the change of the
compliance on the vertical axis.
0.94 ml
0.81 ml
- 37 daPa
32 daPa
ml
3
2
1
0
-600
-300
300 daPa
Right
Ear Volume
Compliance
Pressure
Gradient
0.94 ml
0.81 ml
-37 daPa
31 daPa
Tympanogram
Remarks:
Examiner:
Name:
Id No.:
Date:
MAICO
MI 24
100
Reflex
Right
Ipsi 500 Hz PASS
ml
0,15
0,10
0,05
0
dBHL
s
100
Ipsi 1000 Hz PASS
ml
0,15
0,10
0,05
0
dBHL
s
100
Ipsi 2000 Hz PASS
ml
0,15
0,10
0,05
0
dBHL
s
95
Ipsi 4000 Hz PASS
ml
0,15
0,10
0,05
0
dBHL
s
Figure 24
Printout of a normal tympanogram
Summary of Contents for MI 24
Page 5: ...Operating Instructions MI 24 MI 24h 3 NOTE 1 02 mm H2O 1 0 daPa...
Page 8: ...Operating Instructions MI 24 MI 24h 6...
Page 11: ...Operating Instructions MI 24 MI 24h 9 24...
Page 47: ...Operating Instructions MI 24 MI 24h 45 Tested by Date...
Page 49: ...Operating Instructions MI 24 MI 24h 47...
Page 52: ...Operating Instructions MI 24 MI 24h 50...
Page 53: ...1162 0321 REV E 07 07...