26
Maxuum III Digital Electronic Indicator 2239028 Rev D
segment. Enhanced functions using the FAN display (MIN, MAX
and TIR) require the user to depress the RESET button after
obtaining an on-scale reading. In Enhanced mode, the indicator
reacts like a traditional Maxum Plus, digital readings are
restored with any on-scale reading.
Figure 19 - Indicator shown in Over-Range condition
4.1.3 Standard Mode
This mode operates according to the traditional Maxum unit.
This is the normal static measuring mode and is indicated by a
normal static Fan display and no tolerance arrows when within
tolerance. If out of tolerance one of the arrows will be flashing.
4.1.4 Initial Setup - Standard/TSP Mode
This mode shows the actual position of the spindle in the Indi-
cator's measuring range. When True Spindle Position is acti-
vated, the Indicator's analog display changes to the alternate
mode, i.e. a normal Fan display changes to a Single Element dis-
play and vice-versa.
Use a ballpoint pen or similar pointed instrument to activate
switches. Do not use a pencil; graphite dust may affect button
switches.
1.
Mount the Indicator or Digital Transducer in the gage.
2.
Place the Nominal or Zero Master in the gage.
3.
Press and release the True Spindle Position (TSP)
button. Unit must be on-scale.
4.
Mechanically adjust the Indicator or Digital Transducer
until reading is at or near zero.
Securely lock the Indicator or Digital Transducer in position mak-
ing sure reading is still at or near zero.