Maxuum III Digital Electronic Indicator 2239028 Rev D
4.2.1.1 How to change between the Standard Mode
and the Enhanced Mode:
NOTE: The following steps require the unit to be on-scale. On-
scale means that a digital reading is viewed on the digital dis-
play.
1.
When the True Spindle Position (TSP)
button is
pressed and released, the indicator displays the True
Spindle Position in the Standard Mode. (Section, ’4.1.4
Initial Setup - Standard/TSP Mode’ on page 26)
2.
To enter the Enhanced/Actual Mode, press and hold the
True Spindle Position (TSP)
button (>2 seconds)
until the two tolerance arrows begin to flash
on the display. Readings are static in this mode.
3.
To toggle back to the Standard/TSP Mode press and
hold the True Spindle Position (TSP)
button
(>2
seconds) until the two tolerance arrows disappear from
the display.
4.2.2 Enhanced/Dynamic Mode
The unit has to be in the Enhanced/Actual Mode in order to
enter into the full Enhanced Mode. While in Enhanced/Actual a
press and release of the min, TIR, Hold, or Max button enters
into its corresponding selection. The dynamic functions are
described below.
4.2.2.1 Minimum Reading
Displays the smallest measurement only.
Pressing the Min button selects the Min Mode. The left toler-
ance limit arrow of the display is on and the fan be reads from
full scale MAX to the MIN value. Use the Reset button to clear
the displayed Minimum value and start a new min measure-
ment.
NOTE: (When using this and the following functions, spindle
location is indicated by a single moving segment within the
Fan display.)
4.2.2.2 Maximum Reading
Displays the largest measurement only.
Pressing the MAX button selects the MAX mode, the right tol-
erance limit arrow of the display will be on and the fan will be