Load Cell Zero and Full Weight Calibration
The WSB has a set of load cells for weighing the weigh bin. The load cells must be calibrated periodically with zero
and full weight calibrations for proper operation and accurate weight reporting. The zero calibration routine should be
performed routinely to tare the load cells back to zero or near zero when necessary. The Full weight calibration should
be performed periodically to set the calibration slope. Full Weight Recalibration of the load cells is recommended
every 6 months especially if totals are collected from the blender. The following instructions outline how to calibration
the weigh bin load cells, identified as the load cell assembly that weighs the removable weigh bin.
Before performing a Zero Weight Calibration or a Zero/Full Weight Calibration, verify the following:
BE SURE
the weigh bin is EMPTY.
BE SURE
the load cell cable is plugged into the side of the controller.
BE SURE
the weigh bin is resting on the load cells freely and does not make contact with anything around it.
BE SURE
the air line to the
weigh bin’s dump valve is connected as it would be during normal operation.
BE SURE
The load cells and bin are not jammed in any way. To test for this, see that a light touch on the bin
causes the display to change. When the pressure is removed the display must return to exactly
where it was, plus or minus 1 gram. If this does not happen, bin is not entirely free to move. Check
EVERYTHING around the bin.
Zero Weight Calibration
Press
Display will prompt for a password.
(default: 22222)
Then
press:
Press
Blender Configuration
Display will show menu options
Press
Calibration Routines
Display will show menu options
Press
Calibrate Load Cells
Press
ZERO
Display will say: Visually confirm that the weigh bin
is empty and then press:
YES
(if weigh bin is not
empty it can be dumped using DUMP BIN button).
After pressing ZERO, a gram weight of
0
(+/- a few
grams) should be displayed.
The ZERO calibration of the load cells is complete.
If a FULL WEIGHT calibration is to be performed next, continue to FULL WEIGHT
CALIBRATION on the next page. Otherwise, if only the ZERO calibration is to be
performed at this time, exit calibration mode.
Press
To go back to the Home screen.
FULL weight calibration may also be done at this time.
FULL Weight Calibration, see next page.
Summary of Contents for Weigh Scale Blender 4088
Page 6: ...6 EC Declaration of Conformity ...
Page 10: ...10 Blender Parts Key ...
Page 13: ...13 ...
Page 104: ...104 Technical Drawings ...
Page 105: ...105 ...
Page 106: ...106 ...
Page 107: ...107 ...
Page 108: ...108 ...
Page 109: ...109 ...
Page 110: ...110 ...
Page 111: ...111 ...
Page 112: ...112 ...
Page 113: ...113 ...
Page 114: ...114 ...
Page 115: ...115 ...
Page 116: ...116 FCA INSTALLATION DIAGRAM ...
Page 117: ...117 WSB 4088 IO wiring diagram 120V ...
Page 118: ...118 D WSB E0001 Electrical Diagram WSB Export Single Phase ...
Page 119: ...119 D WSB E0002 Electrical Diagram WSB Export 3 Phase ...
Page 120: ...120 D WSB E0003 Electrical Diagram WSB Export 1800 3 Phase ...
Page 121: ...121 D WSB E0004 Electrical Diagram WSB Export Maxibatch ...
Page 123: ...123 D WSB P0001 Pneumatic Diagram WSB with Non Removable Hoppers ...
Page 124: ...124 D WSB P0002 Pneumatic Diagram WSB with Removable Hoppers ...
Page 125: ...125 ...
Page 143: ...143 Flexbus Lite Wiring Diagrams ...
Page 144: ...144 ...
Page 145: ...145 ...
Page 146: ...146 Flexbus Lite Component Map ...