Magtrol Model 6530 Power Analyzer
Appendix A: Schematics
appendices
Magtrol 7500 Series Power Analyzer
Appendix B: Compatible Command Sets
b.2
data output Commands
Command Code
Function
Explanation
OAm1,m2<terminator> Requests amp measurement
value.
"m1" indicates the input phase.
Values for
m1
are:
0 = sum of amps phases
1 = phase 1
2 = phase 2
3 = phase 3
"m2" indicates the measurement value.
Values for
m2
are:
0 = normal (RMS/DC) depending on
measurement mode setting)
1 = peak (instantaneous)
2 = crest factor
3 = peak hold/inrush (RMS/DC)
OVm1,m2<terminator> Requests volt measurement
value.
"m1" indicates the input phase.
Values for
m1
are:
0 = average of voltage phases
(wiring mode dependent)
1 = phase 1
2 = phase 2
3 = phase 3
"m2" indicates the measurement value.
Values for
m2
are:
0 = normal (RMS/DC) depending on
measurement mode setting)
1 = peak (instantaneous)
2 = crest factor
3 = peak hold/inrush (RMS/DC)
OWm1,m2<terminator> Requests watt measurement.
"m1" indicates the input phase.
Values for
m1
are:
0 = sum of watt phases (wiring mode
dependent)
1 = phase 1
2 = phase 2
3 = phase 3
"m2" indicates the measurement value.
Values for
m2
are:
0 = watts
1 = VA
2 = power factor
3 = watts peak hold
OF<terminator>
Requests output of frequency.
---
OEm<terminator>
Requests phase measurement
values (amps, volts, watts).
"m" indicates the input phase.
Values for
m
are:
0 = sum of phases (wiring mode
dependent)
1 = phase 1
2 = phase 2
3 = phase 3
OAVE <terminator>
Retrieves displayed values of
current, voltage and power.
Returns data in the same
format as the OE command.
Displays values in average mode only
OT<terminator>
Requests all measurement
values.
---