Magtrol Model 6530 Power Analyzer
Appendix A: Schematics
appendices
Magtrol 7500 Series Power Analyzer
Appendix B: Compatible Command Sets
Command Code
function
explanation
RVm1,m2<terminator> Sets voltage range.
"m1" indicates the input phase.
Values for
m1
are:
0 = all phases
1 = phase 1
2 = phase 2
3 = phase 3
"m2" indicates voltage range.
Values for m2 are:
0 = 600 V
1 = 300 V
2 = 150 V
3 = 30 V
SAm1,m2<terminator> Sets the amps scaling constant
(current transformer).
"m1" indicates the input phase.
Values for
m1
are:
0 = all phases
1 = phase 1
2 = phase 2
3 = phase 3
"m2" indicates the amps scaling
constant in A/A and must be set
within the following range: 0.01 < m2
< 10000. When m2 = 0, the amps
scaling mode will be cleared.
SVm1,m2<terminator> Sets the voltage scaling
constant (potential
transformer).
"m1" indicates the input phase.
Values for
m1
are:
0 = all phases
1 = phase 1
2 = phase 2
3 = phase 3
"m2" indicates the voltage scaling
constant in V/V and must be set
within the following range: 0.01 < m2
< 10000. When m2 = 0, the voltage
scaling mode will be cleared.
SEm1,m2<terminator> Sets the external current
sensor scaling constant.
"m1" indicates the input phase.
Values for
m1
are:
0 = all phases
1 = phase 1
2 = phase 2
3 = phase 3
"m2" indicates the external current
sensor scaling constant in A/mV and
must be set within the following range:
0.0001 < m2 < 99999.
note:
External sensor voltage input
values are divided by this value to give
output in amps in external sensor input
current ranges.
WMm<terminator>
Sets the wiring mode for sum
measurement values.
"m" indicates the wiring mode.
Values for
m
are:
0 = 1-Phase, 2-Wire
1 = 1-Phase, 3-Wire
2 = 3-Phase, 3-Wire
3 = 3-Phase, 4-Wire
4 = 3-Volt, 3-Amp