![Magnescale MG80-PN Operating Manual Download Page 51](http://html1.mh-extra.com/html/magnescale/mg80-pn/mg80-pn_operating-manual_3565527051.webp)
50
(22)
Master preset value setting command
Offset
Label
Setting value (Hex)
0
INC
Please enter an arbitrary value different from the value previously
sent.
1
CMD
0x19
2
RSV1
0x00
3
RSV2
0x00
4
DATA1
※
ASCII code
Axis number
Setting value
(Hex)
‘
1
’~’
10
’
0x30
~
0x39
‘
11
’~’
16
’
0x41
~
0x46
5
~
8
DATA2
~
DATA5
Master preset value
※
Integer type 8 digit 4 Byte Value (Hex) (the last 4 digits are
decimals)
(Example)
+
0.1
μ
m
→
+1000(0x3E8)
-
0.1
μ
m
→
-1000(0xFC18)
11
~
15
DATA6
~
DATA11
Dummy
0x00
(23)
Master preset value acquisition command
Offset
Label
Setting value (Hex)
0
INC
Please enter an arbitrary value different from the value previously
sent.
1
CMD
0x1A
2
RSV1
0x00
3
RSV2
0x00
4
DATA1
※
ASCII code
Axis number
Setting value (Hex)
‘
1
’~’
10
’
0x30
~
0x39
‘
11
’~’
16
’
0x41
~
0x46
5
~
15
DATA2
~
DATA11
Dummy
0x00