2
(E)
DZ174-010/-025/-050
5. Mounting the holder
Mount the holder with the supplied M5
✕
16 screws, using 2-ø5.5 mm (0.22" dia.) holes. (Refer to Fig. 2)
Adjust the probe position so it is perpendicular to the measuring surface and parallel to the moving direction
within a tolerance of 0.3 mm/100 mm (0.01"/3.9"). If necessary, mount an arbor to the hole of the mounting
holder.
Fig. 2
Mounting the holder
6. Mounting the probe
1.
Remove the feeler from the probe, and screw it onto the supplied coupler (Fig. 3).
Be sure to apply screw lock or other adhesive to prevent loosening of the feeler.
Tighten the feeler by hand. Do not use pliers or other tools.
Fig. 3
Assembly of the feeler
2.
Mount and loosely secure the probe in the order indicated by the numbers. (Fig. 4)
Fig. 4
Mounting procedure of the probe
2 hex. socket-head
cap screws
M5
✕
16 (supplied)
Holder
2-ø5.5 mm
(0.22" dia.) holes
0.3 mm/100 mm
(0.01"/3.9") or less
0.3 mm/100 mm
(0.01"/3.9") or less
Moving direction
Arbor
Feeler
Coupler (supplied)
Screw in.
After assembly
Apply screw lock.
Mount the probe in the order indicated by the numbers.
q
Loosen the screw.
w
Insert the probe.
e
Surfaces a and b should be roughly even.
r
Tighten the screw to loosely secure the probe.
a
b