5.3.3. The averaging process
The analyser is capable to display all results, measured or calculated, either as
current value or as averaged value. The averaging time can be selected in a range of 10
seconds to 60 minutes. If a time of 1 min was set, the averaged results over that time
would be shown on the display. During the start of the averaging process a row of strokes
appears on the screen, since there are no results to display
.
If measurement parameters are changed during the averaging process, the aver-
aging cycle has to be re-started.
In case the data storage is activated it is stopped as well.
The following incidents cause an interruption of the averaging cycle.
•
change of the
averaging time
•
change of the
reference oxygen
•
change of the
fuel
•
switching the instrument to the
STANDBY
mode