5.5. Soot (smoke) Measurement
To perform a soot test it is essential to prevent the flue gas from condensing on
the filter. Therefore it is necessary to use a heated probe holder. It is only possible to per -
form a soot test if the instrument is connected to the mains.
The soot test is done by means of the Bacharach comparison method.
-
select the main MENU
and
-
move the cursor to
SOOT TEST
-
select the option, the pump is switched off and the
following screen appears:
Insert the soot filter in the slit of the heated probe holder, and place the probe in
the stack.
-
starts the soot measurement. The pump is switched
on and sucks in 60 s (according to. TÜV) the re-
quired volume of 1.63 +/-0.07 l