www.madur.com
Manual
GA-21Plus
5.2.4. Initial calibration
After the instrument is switched on it carries out an initial calibration for 2 minutes.
Warning:
Make sure the probe is not in the stack during the initial calib
ration
Initial calibration is of basic importance for measurement correctness and it must
not be interrupted before it is finished automatically.
During this time the Oxygen sensor is calibrated to 20,95% in ambient air, and
the other sensors (CO, NO,...) will be zeroed.
23