a scan, the spectrum recorded is shown in the detail view. Moreover, in the case of a color
measurement and a scan, the respective analysis menus can be reached by pressing
or
.
The IQC memory backs up all measurement results that are generated via the IQC menu. These
are results of internal quality control which are stored separately from the general measurement
results.
UV/VIS II
Open Mode, 01.03.2016, 10:57
17-09-2014.: 10:46 Standardmessung
Bestanden
17-09-2014.: 10:46 Standardmessung
Bestanden
17-09-2014.: 10:46 Standardmessung
Nicht bestanden
17-09-2014.: 10:46 Standardmessung
Bestanden
17-09-2014.: 10:46 Standardmessung
Bestanden
17-09-2014.: 10:46 Standardmessung
Nicht bestanden
Anfangsdatum
Enddatum
Methode
Benutzer
Probeort
Probetyp
Probenehmer
Anwender
IQK Typ
Fig. 18: IQC memory
The IPC memory is accessed using the
icon
in the task bar. Pressing
followed by
leads
to the IQC memory. The IQC card 4, among other
things, is created from the results stored here.
Analogously to the measured value memory,
the IQC memory can be printed, selected, or
exported. To view detailed information, the same
rules apply as in the case of the measured value
memory.
Cleaning
DANGER
Risk of electrical shock: The device must be switched off and disconnected from the power
supply during all cleaning work.
WARNING
Risk of fire: Do not use any flammable cleaning agents or organic solvents to clean
the device
DANGER
Hazard due to the development of gases containing chlorine: Chlorine compounds can
develop lethal gases under the influence of UV radiation. For cleaning, do not use any
cleaning agents containing chlorine.
NOTICE
Do not clean the device with acetone or similar products.
NOTICE
Ensure that there is no cuvette in the cuvette shaft.
1. Wipe the outside of the device with a damp cloth. Use water or a mild cleaning agent for this
purpose. Then dry the device with a soft cloth.
2. If necessary, clean or dry the touch screen with a soft, lint‑free cotton cloth.
3. If necessary, clean or dry the cuvette shaft with a soft, lint‑free cotton cloth.
Details on cleaning and maintaining the device can be found in the complete operating
instructions of the device.