3
10°
2
4. Tighten the two screws (4) counter clockwise (view from below) with a wrench.
4
GMP 5000, GMP 10 000, GMP 20 000 and GMP 30 000
1. Install the drive unit into the tank plate. Align the four holes in the drive unit to the
threaded holes in the tank plate.
2. Screw the four bolts into the tank plate and tighten them with a wrench.
Electrical Installation
• When working on live drive controllers, prevent accidents by observing applicable
national regulations.
• All electric installations, including the variable frequency drive, must be performed
by an authorized electrician.
• The mixer must be operated with a variable frequency drive to function properly.
• Drive units require the following speed control device:
3-phase AC motor Frequency converter
1-phase DC motor Thyristor controller
Before installation
• Ensure that all incoming power is equipped with an emergency stop and an on/off
switch that can be locked in both positions.
• Ensure that the motor cables are shielded (to avoid electrical disturbance).
!
!
14 www.millipore.com