9
Under the counting version if intend to change the
counting sample units, the new procedures are following :
a.
Take away any load from the platform.
b.
Push the " function A Button " ( 3-3, Fig. 1 ) once a
while. LCD will show no. of sample units from 10, 20,
50, 100 in sequence & changed per second.
c.
Put the counting sample units ( For example 10 PCs,
20 PCs, 50 PCs or 100 PCs ) on the platform. Until the
display reach the no. of sample units exactly, for
example " 50 ", push the " function A Button " ( 3-3,
Fig. 1 ) once a while again. The display will flash
twice and freeze the set counting no., then take away
the sample units. Now the balance is ready for counting
& the sample counting units already change to new
value.
Consideration
:
@
Under the counting function, 1 PC weight
should large than 0.10 g, other wise the
result of counting no. may existing some
error.
@
The accuracy for the " counting no. " result
are always depend on the no. of " sample
units " that key in. We strong recommend
that it can take more sample units as
possible, for example 50 PCs or more, then
will get the best accuracy.
4)
a. Take away the counting sample units away from
the platform, then put the new lot of units that
intend to be counted on the platform.
b. The display will show the counting no. of the new
" sample units ".
Change to the new counting sample units