Step 3 : 

Press the " CAL. Button " ( 3-3, Fig. 1 ) once again, 

display value of " 400.00 g " will flash 6 times, then 

the calibration procedures are finished completely & 

balance  is ready for the measurement accurately. 

 

3)

 

Calibration under the 200 gram standard weight 

a)

 

Power on the scale until the display show zero value. 

Apply the 200 gram standard to the center of the 

platform gently. If the scale is not accurate, then may 

show the weight value around 200 g ( for example 

199.94 g , 200.05 g. ... ) 

 

b)

 

Step 1 : 

Press the " CAL. Button " ( 3-3, Fig. 1 ) once, 

display will show the value of " 200.00 g " 

Step 2 : 

Press the " CAL Button " ( 3-3, Fig. 1 ) once again, 

display value of " 200.00 g " will flash 6 times, then the 

calibration procedures are finished completely & balance 

is ready for the measurement accurately. 

 

Consideration

 

of

 

Self-Calibration

 

:

 

*

 

The

 internal Self-Calibration value is saved to the 

internal EEPROM circuit permanently. Once power off then 

power on, the calibration value will be not disappeared . 

*

 

After make the self calibration, measure the 200 g or 400 g 

standard weight again, if the display value existing few 

counts deviation, it is normal. However within few hours, 

the balance will be kept under a high accuracy condition. 

Those accuracy will be even better than the published 

specification ( refer page 2 ). 

*

 

Under the status of " oz " unit, still can use the weight of 

200 g or 400 g to make the calibration. However the " g " 

marker will disappear. 

Summary of Contents for GM-610P

Page 1: ...E marks a step forward for you into the field of precision measurement Although this DIGITAL BALANCE is a complex and delicate instrument its durable structure developed Please read the following inst...

Page 2: ...B Button 3 3 5 Display 3 3 6 Power Switch 3 3 7 Rubber Pads 3 3 8 Battery Cover Compartment 3 3 9 DC 9V Receptacle 3 3 10 RS232 Output Socket 3 3 11 Unit Switch 3 4 PLATFORM INSTALLATION 4 5 WEIGHT ME...

Page 3: ...cuit permanently even power off High precision counting scale function RS 232 computer interface Accept battery or AC DC adapter power source two way power supply Built in level bubble the adjustable...

Page 4: ...y Data Output RS 232 USB PC computer interface Connect the optional RS232 cable UPCB 02 will get the RS232 plug Connect the optional USB cable USB 01 will get the USB plug Operating Temperature 0 to 5...

Page 5: ...3 Side view Bottom view 3 FRONT PANEL DESCRIPTION Fig 1 Back view Front view...

Page 6: ...3 CAL Button function A Button 3 4 Tare Button function B Button 3 5 Display 3 6 Power Switch 3 7 Rubber Pads 3 8 Battery Cover Compartment 3 9 DC 9V Receptacle 3 10 RS232 Output Socket 3 11 Unit Swi...

Page 7: ...Fig 1 The scale are set to the g unit typically Use the type screw driver to slide the Unit Switch Slide the Unit Switch to the right position will set to gram unit the right down corner of LCD will s...

Page 8: ...en tare function should be executed The max tare capacity is approx 600 g The Tare marker will disappear until power off and power on again 6 SELF CALIBRATION 1 Please prepare the following optional S...

Page 9: ...once again display value of 200 00 g will flash 6 times then the calibration procedures are finished completely balance is ready for the measurement accurately Consideration of Self Calibration The in...

Page 10: ...r away from function A Button 3 a The bottom left corner of LCD will show the COUNT marker at the same time the LCD will show no of sample units from 10 20 50 100 in sequence changed per second b Put...

Page 11: ...set counting no then take away the sample units Now the balance is ready for counting the sample counting units already change to new value Consideration Under the counting function 1 PC weight shoul...

Page 12: ...x 6 PCs into the battery compartment replace the battery cover 1 The scale will also be operated by the household ACV power source 110 220 240 ACV with a DC 9V AC DC Adapter capacity 300 mA at least 2...

Page 13: ...ta stream will be displayed in the following format Each digit indicate the following status D15 Start Word D14 4 D13 1 D12 D11 Annunciator for Display g 57 oz 58 count 30 D10 0 D9 Decimal Point DP po...

Reviews: