2.8 LumiraDx Lot Calibration File installation
LumiraDx Lot Calibration Files are required to provide the Instrument with the
information needed to perform diagnostic tests. Every manufactured Lot of Test
Strips has a unique Lot Calibration File. This contains information about the test
method, the Lot number, the calibration data and expiration date.
Lot Calibration File installation is performed manually. Locate the RFID reader
by looking for the symbol on the side of the Instrument. To install a new Lot
Calibration File, touch the back of the Test Strip Carton to the
symbol to
install. When the Lot Calibration File is installed the Instrument will sound and a
confirmation message will be displayed.
The Instrument will prompt to install the Lot Calibration File when inserting a new
Test Strip Lot. Once installed, the Instrument will have all the information required
to process the test, and any future tests from the same Lot of Test Strips.
24 - Getting Started
Confirmation
The Instrument will sound and a
confirmation message will be displayed.