2.1 Unpacking
The LumiraDx Instrument package includes the following contents:
1.
LumiraDx Instrument
2.
LumiraDx Power Supply Unit
3.
Platform User Manual
4.
Platform Quick Reference Guide (including passwords for Standalone
operation)
Inspect the Instrument and packaging for damage before use. Report any
damage to Customer Services. Refer to the “Customer Services” chapter of this
Platform User Manual for contact information.
2.2 Powering the Instrument on and off
The Instrument must be fully charged before first use. It will take approximately 2
hours to fully charge the battery using the power supply unit provided.
Place the Instrument on a level, stable surface. Power on by briefly pressing the
power button at the rear of the Instrument. You will hear the Instrument powering
on and the display will be blank for several seconds. When prompted enter user
ID and/or password to login.
The Instrument touch-screen will dim after 2 minutes of inactivity. Tap the touch-
screen to restore the touch-screen brightness.
To power the Instrument off after use, press the power button at the rear of the
Instrument for 2 seconds, and tap the screen message to confirm power off. The
Instrument should be powered off for transportation and when not in use.
2.3 Instrument self-check
The Instrument will perform a self-check when starting up. Ensure that the
Instrument is on a level, stable surface during this process. If the Instrument door
is open, the Instrument will prompt to close the door before performing the self-
check. If a Test Strip is inserted at this time, the Instrument will prompt to remove
the Test Strip and close the door. The self-check will not proceed until the door is
closed.
2 Getting Started
20 - Getting Started