MAP Technology
The System
9
•
Reporter control microspheres (CON2)
•
Luminex xMAP Sheath Fluid
xPONENT 3.1
•
Calibration Kit (CAL)
•
Verification Kit (VER)
•
Luminex xMAP Sheath Fluid
Required
Laboratory
Reagents
•
Household bleach
•
70% isopropanol or 70% ethanol
•
Mild detergent
•
Distilled water
Luminex Software
Luminex software provides complete control of the system and
performs data analysis. Your Luminex 200 system is preloaded with
the Luminex software. However, we supply a software DVD should
you need to reinstall the software.
This software requires a dedicated system. Unauthorized additional
software is prohibited and may result in improper operation of the
system.
Luminex 200
Performance
Specifications
Speed
•
USB communications link for fast data transfer
•
Automatic transfer of assay templates and new reagent
information into the system via a large capacity read/write DVD
•
Installation: < 4 hours
•
System calibration: < 10 minutes
•
System controls: < 10 minutes
•
Barcode reader entry of sample IDs
•
Automatic post-analysis
•
Analyze one 96-well plate/hour depending on manufacturer’s kit
•
Up to 100 xMAP microsphere sets per sample
•
System warmup: 30 minutes. Systems that remain inactive for at
least four hours will require a warm-up to restart the lasers. After
acquiring sample, running system calibrators, running system
controls, and warming up the instrument, the system resets the
four-hour internal clock.