Treatment Screen Description
Lumenis Pulse 120H UM-10012510 Rev. F
Page 56
8. Fiber Status
Area
- Certain Lumenis SIS fiber delivery systems for
the Lumenis Pulse 120H are designed to allow several surgical
treatments, while others are limited to only one treatment. When a
fiber is connected, the system immediately knows:
• How many treatments have been performed with the fiber.
• How many treatments are recommended before you replace the
fiber.
• If all allocated treatments are exhausted and the fiber is expired.
• If the delivery system has any power limitations.
Every time the fiber is connected to the system, the fiber status area
will indicate the fiber mode. There are four fiber modes, color-coded
according to the status:
•
Normal mode
(green) – The fiber is working within operational
limits.
•
Grace mode
(orange) – You are advised to replace the fiber,
because it exceeded recommended usage. However, you can
continue to work. The
Fiber exceeded recommend # of uses. It is
advised to replace the fiber
error message will also appear inside
the notification bar.
•
Fiber expired
(red) – You cannot work with this fiber. The
Fiber
expired
recoverable error message will also appear inside the
notification bar.
•
Unrecognized Fiber
(red)
– You cannot work with this fiber. The
Lumenis SIS fiber not detected
recoverable error message will
also appear inside the notification bar.
By pressing the
icon in the
Fiber Status Area
, additional
information regarding the number of sessions used and the number of
recommended sessions for the fiber in use will be displayed.
NOTE:
For detailed information on the number of treatments each Lumenis fiber
is designed to perform, refer to the instruction guide delivered with the
fiber.