No Laser Emission
Lumenis Pulse 120H UM-10012510 Rev. F
Page 115
No Laser Emission
1.
Replace the delivery system.
2.
Inspect and, if necessary, replace the debris shield.
3.
Contact your local Lumenis service representative.
A Notification Appears on the Control Screen
1.
Clear the message by pressing the
Acknowledge
button and follow the
suggested steps intended to clear the fault.
2.
Resume normal operation; if the same notification appears again, turn
off the laser system for one minute and restart it. Resume normal
operation.
3.
If the same notification appears again, record the error number and
contact your local Lumenis service representative.
Cooling System Notification
The following notification appears on the control screen:
System
overheated
.
1.
Press the
Show Notifications
icon to open a pop-up window that
displays the nature of the notification.
2.
Clear the message by pressing the
Acknowledge
button and pause
operation for one minute to allow the system to cool.
3.
Resume normal operation; if the same notification appears again, turn
off the laser system for five minutes and restart it. Resume normal
operation.
4.
If the same notification appears again, record the error number and
contact your local Lumenis service representative.
Fiber Exceeded Recommend # of Uses. It is Advised to Replace the Fiber
1.
Be aware that the fiber has exceeded the recommended number of
uses.
2.
Replace the optical fiber with a new one.