Accessories
Lumenis Pulse 120H UM-10012510 Rev. F
Page 125
Accessories
Delivery Devices and Accessories
• SlimLine / aa SlimLine SIS 200, 365, 550, 1000
• SlimLine EZ / SlimLine EZ SIS 200, 365, 550
• SlimLine GI SIS 365
• Duotome
• Xpeeda D/S/L
• Slimline 200 D/F/L
• Moses D/F/L 200, 365, 550
• Sterile aspiration tubing
• Non-sterile drainage tubing
• Collection container
• Suction handpiece
Additional Accessories
• Fiber cutting scissors
• Optical fiber strippers
• Cleaving tool
• SlimLine steam sterilization tray
• Safety eyewear
Electrical Requirements
Electrical Utilities
The Lumenis Pulse 120H holmium laser is available in two electrical
configurations:
• Single-phase (200-240 VAC, <46A, 50 or 60 Hz)
• Three-phase (380-415 VAC, <18A per phase, 50 Hz)
Electrical power should be setup according to the model ordered. The
service technician will configure the system during installation for the site
voltage and verify that the installed power plug is compatible with the
receptacle provided by the hospital.