Electromagnetic Immunity
Lumenis Pulse 120H UM-10012510 Rev. F
Page 135
Electromagnetic Immunity
Guidance and manufacturer’s declaration – electromagnetic immunity
The Lumenis Pulse 120H is intended for use in the electromagnetic environment specified below. The cus-
tomer or the user of the Lumenis Pulse 120H should assure that it is used in such an environment.
IMMUNITY test IEC 60601 test level
Compliance level
Electromagnetic environment –
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 6 kV contact ± 8 kV
air
± 6 kV contact ± 8 kV
air
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%.
Electrical fast
transient/burst IEC
61000-4-4
± 2 kV for power
supply lines
± 1 kV for input/output
lines
± 2 kV for power
supply lines
± 1 kV for input/
output lines
Mains power quality should be that
of a typical commercial or hospital
environment.
Surge IEC 61000-
4-5
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
± 1 kV line(s) to
line(s) ± 2 kV line(s)
to earth
Mains power quality should be that
of a typical commercial or hospital
environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines IEC
61000-4-11
<5% UT
(>95% dip in UT)
for 0.5 cycle
40% UT (60% dip in
UT) for 5 cycles
70% UT (30% dip in
UT) for 25 cycles
<5% UT (>95% dip in
UT) for 5 sec.
N/A
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of the
Lumenis Pulse 120H requires
continued operation during power
mains interruptions, it is
recommended that the Lumenis
Pulse 120H be powered from an
uninterruptible power supply.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical
commercial or hospital environment.
NOTE: UT is the a.c. mains voltage prior to application of the test level.