Hospital/Clinic Staff Maintenance
Lumenis Pulse 120H UM-10012510 Rev. F
Page 118
Hospital/Clinic Staff Maintenance
Visual Inspection
The exterior of the system should be inspected once a week to ensure that
there are no loose cable connections and that there is no damage to the
system.
Routine Exterior Cleaning
The external surfaces of the system (console, LCD panel) and the
footswitch should be cleaned when the system is received, and thereafter
as required by clinic protocol.
The outer surfaces of the system may be wiped clean with a soft, lint-free
cloth dipped in 70% isopropyl alcohol, or a hospital-grade disinfectant
solution.
CAUTION:
Do not spray or pour cleaning agents directly on the laser console or
control screen. You may damage the console, screen, and laser system
electronics.
Remote Interlock Check
Laser beam emission is disabled when the remote interlock plug is not
connected or is improperly connected to the rear panel, even if it is not
wired to an actual door interlock. To check this:
1.
Set the system to
Standby
mode.
2.
Unplug the remote interlock plug.
3.
Try to select
Ready
mode. The system should display the following
warning message in the notification bar:
Verify door closed
.
4.
If the system does not display the warning message and remains in
Ready
mode, discontinue use and contact Lumenis Service.