where reproducibility and accuracy are essential. To use power regulation, a desired power reference 
value in milliwatts is entered in the onboard control GUI (Figure 5, Control tab [left]). To activate 
power regulation, simply click the padlock icon next to the reference power value (Figure 5, Control 
Tab (Left), 5). Gray shading of the padlock icon and the reference power value shows that power 
regulation is active for the selected output channel. When power regulation is active, the intensity 
setting for the channel is controlled by the onboard microprocessor, based on feedback from the 
light engine’s reference photodiode array. The microprocessor continuously adjusts the intensity 
setting so that the output power matches the power reference value set by the user.

3.3.4 Control via Serial ports  

RETRA (L) Light Engines have two serial ports, labeled USB and RS-232 (Figure 3), which can be set 
to receive either LEGACY or STANDARD mode commands.  Connection to the computer requires a 
USB-A-to-USB B cable (29-10058) or USB-to-RS-232 cable 
(29-10011).  LEGACY commands are limited to controlling on/o

 

switching and intensity adjustment of individual color channels.  Only 
one of the two ports can be set to LEGACY mode at one time.  The 
STANDARD mode command set gives access to an extensive panel of 
operating status reports and configuration settings in addition to the 
the basic control functions of the LEGACY command set.  A complete 
listing of STANDARD mode commands is provided in 

Lumencor Light 

Engine Command Reference

 (Document number 57-10018).  Note that 

LEGACY and STANDARD mode communications use di

erent serial 

protocols (9600,8,N,1 and 115200,8,N,1 respectively).  Changes to the 
command mode setting for a serial port can be made via the 

Settings 

screen of Web GUI (Figure 5, right).  Changes are applied 
instantaneously and are retained between power cycles.

Select the command mode setting for the serial port that is compatible 
with the light engine device driver in the control software.  This 
selection is typically found under the “Devices” tab.  The COM port 
address assigned by the computer to the light engine USB serial port 
must be correctly registered in control software.

3.3.5 Control Light Engine Control Pod 

1. Connect the RETRA (L) Light Engine to pod using USB A-to-USB B 

cable (29-10058).

2. Open the Web GUI control interface as described in Section 3.3.2.

RETRA (L) Light Engine Manual 57-10025

15

lumencor.com

Figure 6

. BNC breakout cable to 

enable TTL triggering

TTL Connector Pin Definitions

OUTPUT
COLOR

Pin

#

OUTPUT
COLOR

Pin

#

RED

1

BLUE

12

GREEN

2

VIOLET

13

CYAN

3

nIR

14

TEAL

11

YELLOW 15

Gnd

6,7,8,10

V

ihigh

 (min) = 2.0V, V

ihigh

 (max) = 5.5V 

V

ilow

 (min) = 0.0V,  V

ilow

 (max) = 0.8V

Summary of Contents for RETRA

Page 1: ...Lumencor Light Engine Manual RETRA L Light Engine RETRA L Light Engine Manual 57 10025 1 lumencor com...

Page 2: ...ification IEC EN 60825 1 2014 Safety of laser products CE Marking Low Voltage Directive 2014 35 EU EMC Directive 2014 30 EU RoHS Directive 2011 65 EU 2015 863 EU REACH Regulation EC No 1907 2006 EC EU...

Page 3: ...ons and Warnings 3 Installation and Operating Instructions 4 Light Output Characteristics 5 Operational Specifications 6 Routine Maintenance and Trouble Shooting 7 Customer Support 8 Warranty RETRA L...

Page 4: ...ht output and generates a reference signal that is applied to the constituent sources in a feedback loop to maintain constant light output over time The light sources within the RETRA L Light Engine a...

Page 5: ...ssement d clarations qui identifient des situations ou des pratiques susceptibles d entra ner des blessures corporelles Caution Statements identify conditions or practices that could result in damage...

Page 6: ...te s curit dans un chemin optique ferm NE PAS pointer la sortie de lumi re directement sur un mat riau susceptible d tre inflammable ou susceptible de br ler Cela comprend tous les tissus les plastiqu...

Page 7: ...nal control and laser aperture that can be separated by 2m or more from a radiation warning device shall itself be provided with a radiation warning device Do not exceed 2m without providing a radiati...

Page 8: ...ss optical components that could be damaged or misaligned by the shock produced by a drop onto a hard surface Attention NE PAS laisser tomber la source lumineuse Elle contient des composants optiques...

Page 9: ...k jumper Figure 2 6 Quickstart Guide instruction document 85 10049 RETRA L Light Engine Manual 57 10025 9 lumencor com AC Power Cords Region Part Number North America 29 10002 Europe 29 10005 United K...

Page 10: ...in decreased product life and or premature failure The RETRA L Light Engine must be operated in a non condensing environment dew point 10 C with controlled ambient temperature 30 C Thermal overload pr...

Page 11: ...tion Initialization of the onboard computer takes about 30 seconds after the master power switch is turned on When initialization is complete the status indicator display Figure 3 will activate The Ke...

Page 12: ...DISTANCE PEUT ENTRA NER UN COMPORTEMENT ANOMAL ET OU ENDOMMAGER LES PRODUITS WARNING Prior to turning the light output on be sure the output end of the liquid light guide or optical fiber is safely d...

Page 13: ...terminated optical fiber connect the fiber to the output adapter using the integral threaded sleeve 2 To enable light output of the RETRA L Light Engine the external gate jumper control key remote int...

Page 14: ...etting Tab right 10 Standby mode is also marked by an automatic shut o of the main cooling fan Standby mode automatically terminates when the next light output ON command is issued Output power regula...

Page 15: ...e command set gives access to an extensive panel of operating status reports and configuration settings in addition to the the basic control functions of the LEGACY command set A complete listing of S...

Page 16: ...utton next to TTL inputs in the GUI under the Settings tab Figure 5 Settings Tab right 12 or send the standard mode command SET TTLENABLE to the LAN port Input TTL signals can be conveniently addresse...

Page 17: ...ificates of conformance are also recorded in Lumencor s database and copies can be requested by e mail to techsupport lumencor com The request message must include the 4 or 5 digit serial number of th...

Page 18: ...the OFF state No light output in response to source ON command serial or TTL There are 4 interlocks that need to be closed check all of these 1 Control key must be inserted in front panel and turned t...

Page 19: ...ns consistent with the specifications given in Section 5 and observing all the Precautions and Warnings notified in Section 2 This warranty does not extend to light engines that have been subject to m...

Page 20: ...RETRA L Light Engine Manual 57 10025 20...

Reviews: