6. As soon the DC power supply is energized, the master power button (top right) will automatically 

light up.

 

 The light engine automatically starts when the power is connected; there is no need to 

push the master power button

 

.

3

7. Wait 30–45 seconds for the initiation sequence (onboard microprocessor boot-up) to complete.

 

 

Do not press any buttons or insert any plugs during this time.

8. When the initiation sequence completes, “LUMENCOR” will flash on front display panel and then 

be replaced by a display showing the current light engine IP address, the internal temperature 
and the fan status.

 

 At the same time, the fan will come on at HI for about 2 seconds and then 

shut o

 automatically.  The light engine is now ready for use. Prior to turning the light output on, 

be sure the output end of the liquid light guide or optical fiber is safely directed into an enclosed 
optical path (e.g. a beam dump).

Notes: 
1

  Ensure that the distal end of the light guide or optical fiber is safely directed into an enclosed 

optical path (e.g. microscope epilluminator or a beam dump) before turning the light output on.  If 
the light engine is equipped with an output adapter for an SMA-terminated optical fiber, connect the 
fiber to the output adapter using the integral threaded sleeve. 

2

  To enable light output of the RETRA (L) Light Engine the external gate jumper, control key, remote 

interlock jumper, and Liquid light guide must be inserted. 

3

  For subsequent start ups, use the master power button to start or shut down the light engine.  

Shut down can also be accomplished using the “Shut Down” button in the GUI (Figure 5, left).

3.3.3 Ethernet Connection and Control GUI 

The Web GUI provides a quick and easy way to control the light engine, providing the ability to turn 
each source on/o

 (Figure 4, Control Tab [left], 2), adjust the power of each source independently 

from o

 to full power (Figure 5, Control Tab [left], 3).  The GUI also tracks the total on time (Figure 5, 

Settings tab [right], 13) and the power sensor readings (Figure 5, Control tab [left], 4). To set up the 
Web GUI, follow the follow protocol below:

1. Go to Start menu > Control Panel > Network & Internet and/ or Network & Sharing Center

2. Click Change Adapter Settings

3. Right-click on Local Area Connection

4. Click on Properties (in pop-up)

5. Select Internet Protocol Version 4 (TCP/IPv4)

6. Click Properties button

7. Use the following IP addresses:

IP Address: 192.168.201.201

Subnet Mask: 255.255.255.0

Default gate way and DNS Server are OK to leave blank

RETRA (L) Light Engine Manual 57-10025

13

lumencor.com

Summary of Contents for RETRA

Page 1: ...Lumencor Light Engine Manual RETRA L Light Engine RETRA L Light Engine Manual 57 10025 1 lumencor com...

Page 2: ...ification IEC EN 60825 1 2014 Safety of laser products CE Marking Low Voltage Directive 2014 35 EU EMC Directive 2014 30 EU RoHS Directive 2011 65 EU 2015 863 EU REACH Regulation EC No 1907 2006 EC EU...

Page 3: ...ons and Warnings 3 Installation and Operating Instructions 4 Light Output Characteristics 5 Operational Specifications 6 Routine Maintenance and Trouble Shooting 7 Customer Support 8 Warranty RETRA L...

Page 4: ...ht output and generates a reference signal that is applied to the constituent sources in a feedback loop to maintain constant light output over time The light sources within the RETRA L Light Engine a...

Page 5: ...ssement d clarations qui identifient des situations ou des pratiques susceptibles d entra ner des blessures corporelles Caution Statements identify conditions or practices that could result in damage...

Page 6: ...te s curit dans un chemin optique ferm NE PAS pointer la sortie de lumi re directement sur un mat riau susceptible d tre inflammable ou susceptible de br ler Cela comprend tous les tissus les plastiqu...

Page 7: ...nal control and laser aperture that can be separated by 2m or more from a radiation warning device shall itself be provided with a radiation warning device Do not exceed 2m without providing a radiati...

Page 8: ...ss optical components that could be damaged or misaligned by the shock produced by a drop onto a hard surface Attention NE PAS laisser tomber la source lumineuse Elle contient des composants optiques...

Page 9: ...k jumper Figure 2 6 Quickstart Guide instruction document 85 10049 RETRA L Light Engine Manual 57 10025 9 lumencor com AC Power Cords Region Part Number North America 29 10002 Europe 29 10005 United K...

Page 10: ...in decreased product life and or premature failure The RETRA L Light Engine must be operated in a non condensing environment dew point 10 C with controlled ambient temperature 30 C Thermal overload pr...

Page 11: ...tion Initialization of the onboard computer takes about 30 seconds after the master power switch is turned on When initialization is complete the status indicator display Figure 3 will activate The Ke...

Page 12: ...DISTANCE PEUT ENTRA NER UN COMPORTEMENT ANOMAL ET OU ENDOMMAGER LES PRODUITS WARNING Prior to turning the light output on be sure the output end of the liquid light guide or optical fiber is safely d...

Page 13: ...terminated optical fiber connect the fiber to the output adapter using the integral threaded sleeve 2 To enable light output of the RETRA L Light Engine the external gate jumper control key remote int...

Page 14: ...etting Tab right 10 Standby mode is also marked by an automatic shut o of the main cooling fan Standby mode automatically terminates when the next light output ON command is issued Output power regula...

Page 15: ...e command set gives access to an extensive panel of operating status reports and configuration settings in addition to the the basic control functions of the LEGACY command set A complete listing of S...

Page 16: ...utton next to TTL inputs in the GUI under the Settings tab Figure 5 Settings Tab right 12 or send the standard mode command SET TTLENABLE to the LAN port Input TTL signals can be conveniently addresse...

Page 17: ...ificates of conformance are also recorded in Lumencor s database and copies can be requested by e mail to techsupport lumencor com The request message must include the 4 or 5 digit serial number of th...

Page 18: ...the OFF state No light output in response to source ON command serial or TTL There are 4 interlocks that need to be closed check all of these 1 Control key must be inserted in front panel and turned t...

Page 19: ...ns consistent with the specifications given in Section 5 and observing all the Precautions and Warnings notified in Section 2 This warranty does not extend to light engines that have been subject to m...

Page 20: ...RETRA L Light Engine Manual 57 10025 20...

Reviews: