7.5 Installation Matters
Because of the existence of the ultrasonic level meter's emission angle and blind zone, we must not
choose the installation position too close to the edge, so as not to affect the normal operation of the
meter
Angle mort
Maximum liquid level
Air distance
Installation height G
Current level H
Lowest liquid level
When the displayed L value is equal to the air distance
it means the installation is good
Still pipe
installation
Flange mounting
Bracket
installation
Do not make
the probe and
the liquid
surface
non-vertical
Do not install
too close to
thecontainer
Do not install
ontop of
obstructions
during sound
wave
transmission
Do not install
in a vacuum
environment
007
Intelligent Integrated 2 Wire, 3 Wire; Split Wall Hanging, Panel Mounting