Model 2363
Technical Manual
Section 7
Ludlum Measurements, Inc.
Page 7-1
January 2015
RS-232 Interface
Communicating with the Model 2363
The Model 2363 has an RS-232 serial port that can be connected to a PC or
other RS-232 device. The proper communication settings are (
38400,8,N,1
):
38400
Baud
8
data bits
N
o parity
1
stop bit
Commands
The following table shows all the commands used to communicate with the
Model 2363. All commands must be in uppercase letters. The data can be in
either upper or lower case. All set commands should return
˝
OK
˝
+ a carriage
return and line feed.
Read Commands
RA
Read sample number
RC
Read calibration date
RD
Read date
REx
Read
efficiencycal constant
RHx
Read header
RI
Read time constant
RL
Read current location
RP
Read paralyzed time
RR
Read rate meter
Section