Model 2363
Technical Manual
Section 7
Ludlum Measurements, Inc.
Page 7-2
January 2015
RS
Read samples
RT
Read current time constant
R1
Read neutron alarm
R2
Read gamma alarm
R3
Read n gamma alarm
R4
Read n gamma Integrated dose alarm
Set Commands
SCmmddyyyy
Set calibration date
SDmmddyyyy
Set current date
SExyyyy
Set
efficiencycal constant
Shxyyyyyyyyyyyyyyy
Set Header
SIxx
Set time constant
SLxxxxxxxxxx
Set current location
SMx
Sets rate meter dumping
SPxxx
Set paralyzed time
SR
Send reset samples
SThhmm
Set current time
S1xxxxxx
Set neutron Rate meter alarm
S2xxxxxx
Set gamma rate meter alarm
S3xxxxxx
Set n gamma rate meter alarm
S4xxxxxx
Set n gamma integrated dose alarm
Command Descriptions
RA
This command reads the current sample number. The output is six characters
including a [CR] and [LF]. The format is:
0001
[CR][LF]
RC
This command reads the calibration date. During power up the Model 2363
checks the current date against this date. If the current date is past the
calibration date, the message “OUTCAL” is displayed and the Model 2363 is
disabled until the calibration date is changed ahead of the current date. The
output is 10 characters including a [CR] and [LF]. The format is:
MMDDYYYY
12251996[CR][LF]