![Ludlum Measurements 3003 User Manual Download Page 53](http://html.mh-extra.com/html/ludlum-measurements/3003/3003_user-manual_1946965053.webp)
Model 3003 and 3003i Multi-Detector Survey Meter / SCA
Section 4
Ludlum Measurements, Inc.
Page 4-23
September 2020
Hundreds digit (0-9)
Use the LOG button to select the appropriate parameter and use the PRG and
ACK/RST buttons to scroll through the parameter options.
Channel 2 - Icon
- Upon entry into Channel 2 Icon, one of the following
parameters will be selected (flashing).
α
– Alpha icon
β – Beta icon
γ – Gamma icon
n- Neutron icon
0FF – No icon
Use the PRG and ACK/RST buttons to scroll through the available parameter
options.
Page 7 - Unit 1
To begin adjusting a unit parameter, short press the PRG or ACK/RST to scroll
through the available unit parameters. When the appropriate parameter is
selected, short press DETECTOR to select the detector that needs the
parameter adjusted. To begin, short press LOG to begin modifying the
parameter chosen. Finally, extra-long pressing the LOG button will complete
the modification.
Unit 1 - RATE/MAX Units and Minimum Display
- Upon entry into Unit
1 RATE/MAX Units and Minimum Display, one of the following parameters
will be selected (flashing).
Note:
Select Unit first before setting other
parameters.
Rightmost Digit (0-9)
Middle Digit (0-9)
Leftmost digit (0-9)
Multiplier with all digits flashing (adjustable from x.xx µ to x.xx k)
Unit (off, dpm, Bq, Bq/cm, cpm, cps, Rem/h, R/hr, Sv/h)
Use the LOG button to select the appropriate parameter and use the PRG and
ACK/RST buttons to scroll through the parameter options.