Model 3003 and 3003i Multi-Detector Survey Meter / SCA
Section 2
Ludlum Measurements, Inc.
Page 2-10 September 2020
ACK/RST button
:
Short Press
(press and release the button after one beep)
In all modes, the button is used to acknowledge (silence) alarms.
In COUNT mode, if the count is initialized, the button will initiate
the count without logging the results upon completion.
If datalogging in Mode 2 and a log event has been initiated, the
button is used to decrement the location ID.
Long Press
(press and release the button after two beeps)
Currently has no effect.
Extra-Long Press
(press and release the button after three beeps)
In RATE mode, it will reset the currently displayed rate if the user
has permission (permission set from Lumic software).
In MAX mode, it will reset the current MAX value if the user has
permission to also reset the RATE mode.
In COUNT mode, it will reset the current count and reinitialize the
instrument to begin the next count. If a count was in progress, it
will not log the results.
In DOSE mode, it will reset the displayed integrated dose if the
user has permission (permission set from Lumic software).
LOG button
:
Short Press
(press and release the button after one beep)
In RATE mode, it will log the current rate reading if datalogging is
enabled.
In MAX mode, it will log the current MAX value if datalogging is
enabled.
In COUNT mode, it will:
o
If COUNT mode is ready: Initiate a count. Upon
completion, the result will be logged if datalogging is
enabled.
o
If COUNT mode is in progress: Button has no effect.
o
If COUNT mode is complete: Button will log the current
result.
In DOSE mode, it will log the current DOSE value if datalogging is
enabled.
If datalogging in Mode 2 and a log event has been initiated, the
button is used to accept the current location ID.
Long Press
(press and release the button after two beeps)