Signals concerning the condition of the measurement system: a slow flashing followed by one or
more short flashes depending on the type of condition:
o
One or more measurements in error: one flash.
o
Recalibration need (factory configurable relatively to operating time and type of measure-
ment cells): two flashes.
•
Signals regarding radio communication condition: two slow flashes followed by one or more short
flashes depending on the type of condition:
o
No radio connectivity: one flash.
o
Connectivity present, being networked: two flashes
•
System operating status alerts: three slow flashes followed by one or more short flashes depending
on the type of condition:
o
No detected functional abnormalities: no signalling.
o
Data memory depleted due to persistent lack of connectivity: one flash.
o
Incorrect configuration: two flashes.
o
Internal anomaly not defined: three flashes.
The alarm condition signalling is indicated by red+blue coloration and is defined as follows:
•
No detected functional abnormalities: no signalling.
•
Alarm detected: a short blink followed by one or more flashes depending on the alarm logic index
configured in the sensor.