Hygiene treatment
31
EN
4. Push the click adapter of the breathing tube into the
device outlet port or if necessary, onto the red drying
adapter.
5. To start the drying process, press the Menu key until
dr0:30
appears in the display. Press
the Softstart key to start the drying process.
Alternatively,
press the On/Off key and the Softstart key simultaneously.
The therapy device now switches on and dries the breathing tube (time: 30 minutes).
The remaining time is shown in the display. Following the end of the drying process,
the therapy device switches off automatically.
– You can interrupt the process at any time by pressing the On/Off key.
– If the breathing tube still has damp spots after drying, start the drying process again.
6. Remove the drying adapter if necessary.
6.3 Clean housing
Warning! Risk of electric shock!
•
Before cleaning, it is essential to disconnect the power cord from the
power supply connection and to remove the plug from the socket.
•
Ensure that no liquids penetrate the therapy device. Never immerse the
therapy device in disinfectants or other liquids, otherwise this may lead
to damage to the therapy device and thus a hazard to users and patients.
Caution!
Ensure that no liquids penetrate the therapy device via the pressure measurement
connection, otherwise there may be malfunctions.
1. Wipe down the therapy device and the power cord with a soft, damp cloth.
2. Take off the filter compartment lid.
3. Clean the filter compartment lid under running water until there are no residues. Then
dry it carefully.
The therapy device must be completely dry before it is started up.