Operation
23
EN
3. To exit the Patient menu, press the On/Off/ESC key.
Alternatively,
if you press no key for 30 seconds, the Patient menu closes automatically.
5.5 Softstart setting
To facilitate falling asleep, specialist staff may set a pressure which deviates from the opti-
mum therapy pressure. This pressure is administered only for a certain period of time.
If the Softstart function on your therapy device is enabled, this time can be selected in
5-minute increments up to a maximum of 45 minutes.
Set time
1. Call up the Patient menu and scroll to the setting for time.
Alternatively,
with the therapy device switched on: keep the Softstart key depressed until the current
setting appears.
2. To change the time, press the Softstart key (
+
) or the Humidifier key (
-
) several times until
the desired time is displayed.
3. To save the setting and exit the menu, press the On/Off/ESC key or wait until the menu
closes automatically.
Softstart On/Off
Status:
therapy device is switched on.
– To switch Softstart on or off, press the Softstart key.
With Softstart switched on, the remaining Softstart time is shown in the display.
5.6 Respiratory air humidifier
Use of a respiratory air humidifier prevents the patient’s upper respiratory tract drying out
during therapy.
Humidifier stage can be set in six stages (1 = low heat output to 6 = maximum heat output).
The most favorable setting for you depends on room temperature and humidity, so the
P
test
16.0
Pressure for mask test: 16 hPa
P
test
20.0
Pressure for mask test: 20 hPa