18
Device setup
EN
Oxygen supply
Warning!
For safety reasons (risk of fire), it is not permitted to supply oxygen directly to the
breathing tube or to the breathing mask without a special safety device. If oxygen
supply valve WM 24042 is used, up to 4 l/min oxygen can be supplied to the breath-
ing mask.
It is essential to follow the safety instructions for handling oxygen as well as the
instructions for use for the oxygen valve and the oxygen device used.
An oxygen concentrator, liquid oxygen or an oxygen cylin-
der with the appropriate pressure reducer can also be
used. This use must be prescribed by the doctor supervis-
ing treatment.