22
Operation
EN
Activate/deactivate Auto On/Off
Status:
therapy device is switched off.
1. Call up the Patient menu and scroll to the setting for Auto On/Off.
2. Press the Softstart key (
+
) or the Humidifier key (
-
) until the desired setting
appears in the
display.
3. To save the setting and exit the Patient menu, press the On/Off/ESC key or wait until the
menu closes automatically.
5.4 Mask test
If the mask test is activated, a higher pressure is output for 30 seconds after the device is
switched on. This gives you the option of checking that the mask is properly located at the
start of therapy and correcting it.
Activate/deactivate mask test
Status:
therapy device is switched off.
1. Call up the Patient menu and scroll to the setting for Mask test.
2. Press the Softstart key (+) or the Humidifier key (-) until the desired setting appears in the
display.
Auto
0n
Auto On/Off activated
As soon as you breathe into the mask (pressure > 0.8 hPa), the therapy
device switches on automatically.
When the mask is removed, the therapy device switches off automat-
ically after 5 seconds.
Auto
0FF
Auto On/Off deactivated
The therapy device can only be switched on by pressing the On/Off
key.
If the therapy device is not used for 15 minutes, it switches off auto-
matically.
P
test
--
Mask test deactivated
P
test
8.0
Pressure for mask test: 8 hPa
P
test
12.0
Pressure for mask test: 12 hPa