Hygiene treatment
29
EN
6. Hygiene treatment
Note
•
Follow the instructions regarding the hygiene treatment of the accesso-
ries (exhalation system, breathing mask, headgear/head strap) in the rel-
evant instructions for use.
•
This product may contain disposable items.
Disposable items are in-
tended to be used only once. So use these items only once and do
not
reprocess them. Reprocessing disposable items may impair the function-
ality and safety of the product and lead to unforeseeable reactions as a
result of ageing, embrittlement, wear, thermal load, the effects of chem-
ical processes, etc.
6.1 Intervals
Daily
Clean breathing mask and exhalation system.
See associated
instructions for use
Clean respiratory air humidifier.
See associated
instructions for use
Clean bacteria filter.
Replace particulate filter of bacteria filter after no
more than 24 hours’ use.
See associated
instructions for use
Weekly
Clean coarse dust filter.
Check fine filter. Replace if required.
Check breathing tube. Clean if required.
Wipe down housing.
Wash headgear/head strap.
See associated
instructions for use
Monthly
Replace fine filter (depending on soiling, but after
no more than 250 hours/one month).
Clean breathing tube.
Every 6
months
Replace coarse dust filter
Every 12
months
Replace breathing mask and exhalation system.
Replace breathing tube.
Replace headgear/head strap.