Chapter
3
Device overview
Level sensor
18
09382-03 - 06.05.2021 - LM 2000
Components of the LM 2000 humidifier
(1)
Symbol for hooking/unhooking the water
chamber
(2)
LED light
(3)
Heating plate
(4)
Ventilation slots
(5)
Touchscreen display
(6)
Heating cable socket (red)
(7)
Temperature sensor cable socket (blue)
(8)
Water level sensor
Operating parameters
The control temperature range at the outlet of the
water chamber can be set to between 29°C and
37°C. The control temperature interval at the sensor
near the patient can be set to between 30°C and
40°C. The difference between the control
temperature at the sensor near the patient and at the
output of the water chamber is b1°C and
+4°C.
The graphic representations of the humidifier are
indicative and only serve to facilitate its possible
installation. The choice of ventilation system is the
sole responsibility of the doctor who initiates the
therapy.
The usage of unauthorised accessories will result in damage to health
and property!
The usage of unauthorised accessories may lead to damage to the health
of the patient and the device.
▪
Use authorised accessories only.
WARNING
Level sensor
The humidifier is equipped with an optical detection
system for the maximum and minimum water level in
the water chamber. This system triggers a first level
alarm if the maximum level is exceeded or a second
level alarm if the water level in the water chamber is
insufficient.
You can activate the optical sensor to detect the
minimum water level from the settings menu.