Chapter
7
Alarms
First-level alarms
LM 2000 - 06.05.2021 - 09382-03
45
Table 8: First-level alarms
Icons on
the
display
Alarm
description
Triggering state
Causes and/or remedies
Error
codes
Alarms
without
latching
function
Alarm system
delays
Heating plate
temperature
too high
The heating plate
temperature has
exceed 95°C.
Check for correct insertion of the water chamber.
E0012
Yes
< 5 seconds
Humidifier
overheated
The internal
temperature of the
LM 2000 has
exceeded 70°C
Wait a few minutes before resuming operation.
Check that the permissible operating conditions are observed under
ambient conditions.
Check that the humidifier is correctly positioned and that the ventilation
openings (under and behind the device) are not blocked.
E0014
Yes
< 5 seconds