Chapter
6
Operation
Adult configuration
40
09382-03 - 06.05.2021 - LM 2000
To scroll between the pages of the parameter list,
use the up and down arrows.
Press the [Cancel] button to exit the settings menu
without saving the changes.
Press the button corresponding to the parameter you
want to change.
This takes you to the screen where you can change
the parameter (example: screen saver).
By pressing the [+] and [-] buttons, you can edit the
following parameters:
▪
Selection from 13 languages
▪
Volume
– Start-up music
– Buttons
– Alarms
▪
Activation time of the screen saver
In the Settings menu, you can also enable the
display setting of the current temperature measured
by the sensor at the outlet of the water chamber
([Chamber temperature] button) and enable or
disable the display of the set temperatures in the
water chamber and near the patient ([Set
temperature] button). These temperatures are
displayed in green on the operating screen below the
respective temperatures measured at the outlet of
the water chamber and near the patient.
Use the [Alarm MAX. Water Level in Chamber]
button to disable the maximum level sensor of the
water chamber.
Use the [Alarm MIN. Water Level in Chamber] button
to enable the maximum level sensor of the water
chamber.
The Reset button resets the parameter to the factory
setting.